Several publications have directly linked adverse patient outcomes, including increased morbidity, mortality and inpatient length of stay (LOS), with overcrowding or LOS in the emergency department (ED).1-3 Recent studies have highlighted that reducing the percentage of turnaround time outliers for pathology tests can reduce LOS in the ED,4 even though standard laboratory processes have been mature for a long time.5 Alternative approaches that use point-of-care (POC) testing in the ED can reduce pathology test turnaround times and LOS in the ED,6,7 but this is not always the case.8,9
The pilot trial was a prospective observational analysis of de-identified trial data compared with historical control data. Historical data represented patients attending one hospital ED before implementation of a redesigned pathology process, and trial data represented patients attending the same ED after implementation of the redesigned pathology process (Box 1). Historical and trial data spanned October 2004 to March 2005, thus minimising potential seasonal effects on LOS in the ED, and included the triage category for each patient. The outcome measure was total ED time.10
The multisite study was a long-term retrospective analysis of de-identified data (with no exclusions) representing 132 521 FBC requests for patients attending seven EDs that implemented pathology process changes (coloured specimen transport bags alone, or coloured specimen bags plus blood tubes with a priority indicator) (Box 1). The outcome measures were the total pathology turnaround time (collected-to-validated time) and within-laboratory turnaround time (received-to-validated time) for FBC requests that did not require examination of a blood film and that were validated automatically via a computer algorithm.
COULTER Gen·S System 2 (Beckman Coulter, Fla, USA) and back-up COULTER MAXM;
COULTER Gen·S System 2; and
Two Sysmex XE-2100 (Sysmex, Kobe, Japan).
Statistical analysis was performed using SPSS version 14 (SPSS Australasia, Sydney, NSW). As the data were not normally distributed, non-parametric statistical testing was used. Summary values are expressed as means, or medians with interquartile ranges. In the pilot trial, the Mann–Whitney U test was used to compare LOS between historical control and trial groups and Pearson χ2 analysis was used to compare the percentages of patients in each triage group for trial versus control groups. Differences between Kaplan–Meier plots were measured by log-rank (Mantel–Cox) analysis.
For the pilot trial, analysed data meeting the inclusion criteria represented 789 patients from the historical control group and 730 patients from the trial group. For Category 1 sites of the multisite study, 65 924 FBC requests met the inclusion criterion for data analysis, of which 46 093 (69.9%) met autovalidation rules and 39 097 (59.3%) met autovalidation rules and had a valid time of specimen collection. For Category 2 sites of the multisite study, 66 597 FBC requests met the inclusion criterion for data analysis, of which 47 036 (70.6%) met autovalidation rules and 37 413 (56.2%) met autovalidation rules and had a valid time of specimen collection.
In the pilot trial, the proportion of patients in the trial group for each triage category was not statistically different to that of the control group (P = 0.08; Box 2). The redesigned pathology process was associated with a 29-minute (15.6%) reduction in median LOS in the ED (P < 0.001; Box 3). Considerable reductions in LOS were also revealed in triage categories 2, 3 and 4 (Box 4).
The use of coloured specimen transport bags plus blood tubes with an incorporated priority indicator showed a sustained and highly significant reduction in pathology test result turnaround time for FBC requests for patients attending EDs when compared with FBC requests that used coloured specimen bags alone (with standard blood tubes). A reduction in total pathology turnaround time was revealed for the entire patient population (P < 0.001; Box 5). The percentage improvements for the total laboratory turnaround time and within-laboratory turnaround time were similar (20.1% and 17.8% reductions, respectively; Box 6). Also, these proportional improvements are similar to the 15.6% proportional improvement in ED LOS in the pilot study.
In addition, there was a highly significant reduction in turnaround time outliers when the data were analysed according to timeframes specified in the Australian Council on Healthcare Standards (ACHS) Pathology Indicators (Box 7).12 Again, the percentage improvements for reduction in outliers for total laboratory turnaround times (49.7%) and within-laboratory turnaround times (44.2%) were similar.
The request–test–report cycle for pathology tests is a multistep process, involving up to 20 people for each sample. Accumulation of low rates of human error during the cycle can contribute to significant failure rates. Failure rates for urgent tests commonly requested by EDs can exceed 20%.13
Laboratories provide up to 80% of the information used by clinicians to make important medical decisions.14 However, in many instances, the percentage of laboratory turnaround time outliers has not changed significantly for up to 10 years.15-17 It has been suggested that timeliness of result reporting has not been a major focus in clinical laboratories16 and that, for at least some critical tests, the actual turnaround times fail to meet the expectations of the test provider and test user.16-19
Faster access to pathology results can reduce LOS in EDs and thereby improve clinical care and reduce total cost of care for individual patients.4,7,14,17,20,21
Several approaches have been used to try to improve laboratory services for EDs. Point-of-care testing, phlebotomists dedicated to ED collections, a dedicated satellite laboratory in the ED, and large-scale laboratory automation systems are costly and unlikely to meet the needs of most hospitals and laboratories as cost-effective solutions to poor turnaround times for analysis of samples from the ED.22,23
Pre- and post-analytical aspects of the pathology request–test–report cycle comprise a larger component of the testing process than the analysis. Improvements in laboratory turnaround time parameters have been achieved by the implementation of lean processing initiatives in the pre-analytical aspect of testing.5 These changes may be easier to achieve and more cost-effective than changes to specific analytical aspects of specimen analysis.5,17,18
By extrapolating our data to public EDs in Queensland, about 247 000 ED patients per year have an FBC that is autovalidated. Thus, the use of coloured specimen bags plus blood tubes with a priority indicator, compared with coloured specimen bags alone, would result in a time saving of more than 32 000 hours per year (which would benefit all ED patients), and more than 20 000 additional samples would meet the ACHS Pathology Indicator FBC collected-to-validated time of less than 60 minutes.12
There is an imperative to improve many aspects of the health care system,24 including patient flow in EDs, throughout the world. Rapid access to diagnostic tests is a prerequisite for good clinical outcomes.25 The initiatives described here are simple and cost-effective, and can be readily implemented at any hospital with an on-site laboratory.
2 Triage categories of patients attending an emergency department during a pilot trial of pathology process redesign
* Number of days for which data were collected and analysed. † According to the Australasian Triage Scale.11 |
3 Proportion of patients discharged from the emergency department, by length of stay, during a pilot trial of pathology process redesign*
![]() | |||||||||||||||
4 Patient length of stay (LOS) in the emergency department during a pilot trial of pathology process redesign
5 Proportion of full blood count results validated, by collected-to-validated time, in a multisite study of pathology process redesign
![]() | |||||||||||||||
* Coloured specimen bags alone. † Coloured specimen bags plus blood tubes with a priority indicator. |
6 Mean laboratory turnaround times for full blood count requests from emergency departments in a multisite study of pathology process redesign
Category 2 sites: coloured specimen bags plus blood tubes with a priority indicator |
|||||||||||||||
7 Turnaround time outliers for full blood count (FBC) requests* from emergency departments in a multisite study of pathology process redesign
Category 2 sites: coloured specimen bags plus blood tubes with a priority indicator |
|||||||||||||||
* FBC requests not processed within standard timeframes according to the Australian Council on Healthcare Standards Pathology Indicators, version 3.12 † Received-to-validated time. ‡ Collected-to-validated time. |
Received 10 November 2008, accepted 16 March 2009
Abstract
Objectives: To determine whether redesign of pathology processes, including indicators of sample priority, could reduce patient length of stay (LOS) in an emergency department (ED), and assess the long-term impact of two indicators of sample priority on pathology clinical performance indicators for ED samples.
Design, setting and participants: Two observational studies of de-identified data from standard databases were conducted — a single-site pilot trial of patients attending the ED of one hospital compared with historical controls, and a multisite study of 132 521 full blood count (FBC) requests for patients attending seven EDs that utilised either of two pathology process changes (coloured specimen transport bags alone, or coloured specimen bags plus blood tubes with a priority indicator).
Main outcome measures: LOS in the ED was measured for the pilot trial, and collected-to-validated times for FBCs that fulfilled computer algorithm validation rules were measured for the multisite study.
Results: In the pilot trial, the redesigned pathology process resulted in a 29-minute reduction (15.6%) in the median ED LOS for all patients (P < 0.001) compared with historical controls. In the multisite study, use of coloured specimen bags plus blood tubes with a priority indicator resulted in an 8-minute reduction (20.1%) in mean collected-to-validated times for FBC requests compared with FBC requests that used coloured specimen bags alone (P < 0.001).
Conclusions: Our pilot trial revealed a direct relationship between pathology process design and LOS in the ED, suggesting that redesigned pathology processes can significantly reduce LOS in the ED. Our multisite study showed that collecting samples directly into blood tubes with an incorporated priority indicator reduces pathology test turnaround times. These data suggest that LOS in the ED can be significantly reduced by simple changes to pathology processes, such as collecting samples directly into specimen containers with an incorporated priority indicator.