Hollow-bore needlestick injuries in a tertiary teaching hospital: epidemiology, education and engineering

R Michael Whitby and Mary-Louise McLaws
Med J Aust 2002; 177 (8): 418-422. || doi: 10.5694/j.1326-5377.2002.tb04881.x
Published online: 21 October 2002

Abstract

Objective: To describe the frequency, cause and potential cost of prevention of hollow-bore dirty needlestick injury (NSI) sustained by healthcare workers.

Design and participants: Ten-year prospective surveillance study, 1990–1999, with triennial anonymous questionnaire surveys of nursing staff.

Setting: 800-bed university tertiary referral hospital in Brisbane, Australia.

Main outcome measures: Rates and circumstances of NSI in medical, nursing and non-clinical staff; knowledge of NSI consequences in nurses; and minimum costs of safety devices.

Results: Between 1990 and 1999, there was a significant increase (P < 0.001) in the trend of the reported rate of NSI. Of the 1836 "dirty" NSIs reported, most were sustained in nursing (66.2%) and medical (16.8%) staff, with 62.7% sustained before disposal. Hollow-bore injuries from hypodermic needles (83.3%) and winged butterfly needles (9.8%) were over-represented. Knowledge among nursing staff of some of the risks and outcomes of NSI improved over the decade. A trend (χ2 = 9.89; df = 9; P = 0.0016) with increasing rate of reported injuries in this group was detected. The estimated cost of consumables only, associated with the introduction of self-retracting safety syringes with concomitant elimination of butterfly needles, where practicable, would be about $365 000 per year.

Conclusion: More than one NSI occurs for every two days of hospital operation. Introduction of self-retracting safety syringes and elimination of butterfly needles should reduce the current hollow-bore NSI by more than 70% and almost halve the total incidence of NSI.

The recognition in the 1980s of the potential transmission of bloodborne viruses to healthcare workers through needlestick injury (NSI) led to a universal focus on the prevention of such incidents.1-3 The frequency and nature of injury with dirty hollow-bore needles experienced by healthcare workers in Australian hospital practice were last published almost a decade ago.1,2 More recently, media concern has turned to the dangers of injuries to members of the public due to discarded needles. Although there is intense lay and political concern over community-acquired NSI, these injuries are uncommon and regarded as lower risk than those sustained by healthcare workers.4,5

We describe the rate of reporting and nature of exposure over the past decade from hollow-bore dirty needles in healthcare workers in a tertiary referral hospital. We have also assessed the benefits and consumables cost of widespread introduction of safety devices into the hospital.

Methods

Princess Alexandra Hospital, Brisbane, is an 800-bed university tertiary referral teaching hospital. Prospective data on staff, self-reported to the infection control service since 1987 as part of the hospital's NSI follow-up protocol, detail percutaneous exposures, including "clean" and "dirty" (ie, injury with a used sharp) NSI and body fluid. Initially, these data were collated and analysed manually, and subsequently with software packages: NOSO 3 (1990–1993),6 Q Logic (1994–1995)7 and EpiNet (1996–1999).8 Variables within the database for these periods remained essentially unchanged. We excluded body fluid exposures and "clean" NSI, as they pose a low or nil risk, respectively, of bloodborne virus transmission. Staff have been grouped into three classes: clinical (nursing, medical); those at risk of "downstream" injury (housekeeping, catering, wardsmen, Central Sterilizing Supply Department laundry); and other staff (phlebotomists, paramedical, laboratory).

Standardised, anonymous questionnaire surveys in convenience samples of nursing staff were also undertaken throughout the surveillance period at about three-year intervals to determine both the true reporting rate of dirty NSI in this occupational group and their knowledge of potential disease transmission.

Frequencies, rates and proportions were calculated using SPSS for Windows.9 Epi Info10 was used to calculate 95% confidence intervals around proportions, ?2 for trend and Fisher exact tests for significance. Annual rates of NSI were calculated using the number of full-time-equivalent (FTE) positions for each year as denominator. FTEs were provided by the hospital's finance department, and were calculated for each staff category as the mean of the number of FTE positions at the end of each financial year of the study period. We assumed that significant fluctuations in the number of FTEs due to annual leave or changes in staffing levels did not occur.

Results
Epidemiology

In the 10-year surveillance period, 1836 reported NSI events were classified as "dirty" (Box 1). The reported rate of dirty NSI increased significantly (P < 0.001) over the surveillance period, with 66.2% occurring in nurses, and 16.8% in doctors.

The causal device was recorded in 1478 medical and nursing incidents. Injuries not implicating a hollow-bore needle occurred on 513 (34.7%) occasions; of these, the two major contributors were solid suture needles (158/513; 30.8%) and scalpel blades (68/513; 13.3%). The remainder (965/1478; 65.3%) involved hollow-bore needles.

These 965 hollow-bore NSI included injuries due to intravenous cannulas (47; 4.9%), butterfly needles (95; 9.8%) and hypodermic needles (804; 83.3%). Clinical activities associated with NSI involving hollow-bore needles included intramuscular or subcutaneous injections (201; 20.8%), direct IV injections (185; 19.2%), obtaining venous or arterial blood (133; 13.8%), aspirating or injecting into an IV line (82; 8.5%), and other, unspecified activities (345; 35.8%). Of the injuries involving hollow-bore needles, 287 (29.7%) occurred during use, 320 (33.2%) after use but before disposal, 89 (9.2%) during disposal, and 250 (25.9%) during other or unknown circumstances. Recapping of hollow-bore needles caused 134 injuries, with NSI being 3.44 times more likely (P < 0.001) to be associated with recapping in the first five years of the surveillance period compared with the latter five years.

Of the injuries sustained by all categories of staff, most (62/104; 60%) of the incidents involving butterfly needles occurred after use, but before or during disposal. Of the IV cannula incidents sustained by all staff, 24/56 (43%) occurred in areas of high dependence: 15 in the emergency department and nine in the operating theatres or intensive care unit. No difference was found in the frequency of IV cannula injuries by hospital location over the 10-year period (P = 0.694).

Circumstances associated with NSI in all staff for the entire survey period (Box 2) illustrate that, although downstream injury (ie, injury after disposal in non-clinical staff) contributed to relatively few incidents (11%), more of these were associated with injury during cleaning, reflecting equipment design, than with careless disposal. Only seven injuries were recorded after appropriate disposal, mainly due to overfilling of containers.

One hundred and twenty-seven (6.9%) staff members were exposed to potential infection through NSI from patients with known bloodborne viruses. Five staff suffered injuries associated with HIV-positive patients. Twenty-three, of whom 19 were immune, were exposed to hepatitis B virus (HBV). The incidence of immunity to HBV of staff reporting an NSI increased significantly (trend ?2 = 7.685; df = 1; P = 0.006), from 60.6% (95% CI, 52.0%–68.7%) in 1990 to 89.4% (95% CI, 84.6%–93.2%) by 1999. Since testing of source for hepatitis C began in 1991, 99 staff members have been exposed. No seroconversion to HIV, HBV or HCV infection was recorded.

Education

We have previously published data demonstrating an improvement of NSI incidence and reporting in nursing staff in the late 1980s.11 Regular anonymous surveys from 1987 onwards (Box 3) demonstrated a highly significant (P < 0.001) initial decline in the number of nursing staff who reported sustaining an NSI, although the rate from 1989 onwards has been constant. Similarly, over the period, failure of nursing staff to report NSI significantly fell (P = 0.0015) to a low of 4% in 1992, followed by a constant rate of under-reporting of 14%–24% as the decade progressed.

Responses to the knowledge questionnaire indicated an increasingly held view (?2 = 50.45; df = 4; P < 0.001) that HBV and HIV (?2 = 28.54; df = 4; P = 0.001) are transmitted by NSI but are not fatal diseases. However, the incorrect perception that HCV is a terminal illness increased from 16.8% (95% CI, 11.8%–22.7%) in 1995 (when it was added to the questionnaire) to 25.1% (95% CI, 19.5%–31.5%) in 2001.

Engineering

Protective IV equipment can be classified by the likelihood of transmission of a bloodborne virus from the NSI it is designed to prevent (Box 4). Although needleless IV systems have been widely implemented, price remains a major barrier to introducing other devices. In general, protective designs are sold at a premium compared with "non-safety" alternatives. Based on our hospital in 2000, the cost of consumables for the device and the likely reduction in NSI associated with hospital-wide introduction of these safety devices is illustrated in Box 4. This calculation assumes that, where a device is introduced, it will prevent all the NSI associated with non-protective equipment. Categories are not mutually exclusive (eg, NSI from aspiration from the side-arm of an intravenous line will be eliminated by introduction of a needleless IV system, but also potentially reduced if the syringe used in the procedure incorporates a self-retracting needle).

Discussion

The only similar Australian study published reported a rate of 9.4/100 FTE per year in nursing staff,1 similar to our hospital's mean rate of 8.79/100 FTE per year in nursing staff over the 10-year survey period (1990–1999). Only one overseas study12 is comparable, as it too followed NSI rates in a single institute after multifocused interventions. The researchers assessed NSI rates per FTE per year over an eight-year period (1990–1998).12 Hollow-bore NSI in all healthcare workers, excluding physicians (who were not assessed), declined from 65/100 FTE per year to 16/100 FTE per year. However, even this improved rate of 16/100 FTE per year is more than twofold higher than that reported previously in Australia1 and in our study.

Although non-hollow-bore sharps injuries constitute a significant proportion of incidents (34.7%), interventions to prevent NSI have focused on hollow-bore needles, which were the most frequent source of NSI and pose the greatest risk of bloodborne virus transmission.13 Orientation and reinforcement education sessions were conducted over the entire survey period. Additionally, awareness and "ownership" were facilitated by competitions held to develop educational posters that were subsequently displayed in all clinical areas. Physical interventions included introduction of a recapping device11,14 (1987) to allow safe recapping as an alternative to immediate disposal; positioning of approved "sharps" bins in areas of clinical activity and on trolleys wheeled to the bedside to facilitate immediate needle disposal15,16 (1989); and introduction of a commercial closed disposal system with sealed containers removed from the hospital by contractors (1989). Safety intravenous cannulas were introduced into the emergency department and operating theatre suite in 1994, but not more widely. The hospital has not used a needleless intravenous (IV) system during the past 15 years, except in the operating theatres from 1995 onwards.

Certain areas of improvement have been identified, including reduction in recapping11 and increasing knowledge in nurses of the possible outcome of bloodborne virus infection. However, the NSI rate per year at Princess Alexandra Hospital (PAH) has not declined over the past decade. The overall increase in the total reported rate may be artificial. It might be explained by better reporting of NSI by medical staff (Box 1), whose rate of NSI as an occupational group now exceeds that of nurses. In contrast, repeated anonymous surveys of nurses suggest that their rate of reporting dramatically improved to the early 1990s. Since then, a constant 76%–96% of the true rate has been maintained for the rest of the decade (Box 3).

The reduction in hazard achieved from better equipment design is pivotal to improvement in safety and quality in all areas of healthcare facilities.17,18 While our study has demonstrated an increased knowledge and awareness of the consequences of NSI in nurses, a high endemic level of NSI continues (Box 1). Further improvement in NSI safety is likely to come not from low cost educational programs, but from engineering.19 The benefits of safer devices over behaviour modification have been widely debated,17,20 and culminated in 1999 in the United States with the passage of the Needlestick Injury Safety and Prevention Act 2000.21 This requires the bloodborne virus exposure control plans of US hospitals to "reflect changes in technology that eliminate or reduce the exposure to blood borne pathogens". Multiple safety devices have been developed and variable effects reported.22-26 All the devices are more expensive than those they replace, and thus their widespread introduction carries significant economic cost. The benefits and limitations of such interventions need careful assessment against the background epidemiology of percutaneous exposures within a given healthcare setting.

Our study indicates that 127 (6.9%) of reported dirty NSI at PAH over a decade were associated with a risk of bloodborne virus acquisition. No seroconversion was detected, but, with the low risk of transmission associated with a single exposure,27 this outcome was not unexpected. Most healthcare workers are particularly concerned about exposure to HIV. However, while dependent on the background carriage rate of bloodborne virus in a facility's patient population, our experience indicates that Australian healthcare workers are at greatest risk of HCV infection, a disease for which no proven effective prophylaxis exists.

Our data suggest that the introduction of syringes with self-retracting needles could reduce high-risk NSI in our hospital by up to 62%. These devices incorporate automatic retraction of the needle into the barrel of the syringe, provided the user depresses the plunger to the end of the barrel. As this is the normal action when expelling the contents of a syringe, regardless of purpose, it is likely to be carried out without conscious intent. Butterfly needles have again been associated with significant risk of NSI to healthcare workers.2,28 While protective shielded butterfly needles have been shown to reduce NSI in one trial,29 they require conscious manipulation of a sheath over the needle at the end of a procedure. There is a need for constant educational reinforcement when using these devices,29 and this does not augur well for their long term effectiveness in routine clinical practice. Replacement of non-shielded butterfly needles, where practicable, with self-retracting syringes could further reduce NSI by up to 10%.

The crude estimated extra cost per year at Princess Alexandra Hospital for syringes with self-retracting needles at $0.65 per unit is $365 000. This figure is conservative, not taking into account indirect savings from reduction in loss of staff time associated with NSI assessment and follow-up, decreased consumables for laboratory testing of staff and patients, and the intangible benefits of reduced staff anxiety.30 Our estimate is not intended to represent a formal cost–benefit analysis, but simply the likely funding necessary to implement purchase of consumables.

Our data show that, at one tertiary referral hospital, more than one potentially high-risk NSI occurs for every two days of hospital operation. Hospital administrators have been shown to be tolerant of a finite NSI rate, although the acceptable threshold varies with their awareness and knowledge of the issue.31 It is they who must determine whether the budgetary implications of protective devices are justified by the potential reduction in these exposures. This is not the case in the US, where legislation was introduced without consideration of either its financial impact on individual healthcare facilities, or the differing impacts of varying safety devices on reduction of high-risk exposure. Such a situation should not be allowed to occur in Australia. Introduction of safety devices should not simply be an emotional response to heathcare worker concerns, but must be shown to have proven efficacy and must be accompanied by a cost-containment strategy. Only well-designed clinical trials can provide these data.

Received 26 February 2002, accepted 1 July 2002

  • R Michael Whitby1
  • Mary-Louise McLaws2

  • 1 Infection Management Services, Princess Alexandra Hospital, Brisbane, QLD.
  • 2 School of Public Health and Community Medicine, University of New South Wales, Sydney, NSW.


Correspondence: whitbym@health.qld.gov.au

Acknowledgements: 

We thank all PAH Infection Control Practitioners who have contributed to the data collection over the past 15 years.

Competing interests:

None identified.

  • 1. Mallon DFJ, Shearwood DW, Mallal SA, et al. Exposure to bloodborne infections in healthcare workers. Med J Aust 1992; 157: 592-595.
  • 2. Bowden FJ, Pollett B, Birrell F, et al. Occupational exposure to the human immunodeficiency virus and other bloodborne pathogens: a six year prospective study. Med J Aust 1993; 158: 810-812.
  • 3. Jagger J, Hunt EH, Brand-Elnaggar J, et al. Rates of needlestick injuries caused by various devices in a University hospital. N Engl J Med 1995; 1: 1-11.
  • 4. Philipp R. Community needlestick accident data and trends in environmental quality. Public Health 1993; 107: 363-369.
  • 5. Gerrard JG. Needlestick injuries from syringes discarded in public places. Southern Queensland Network Newsletter 2000; April; 1-4.
  • 6. NOSO 3 [computer program]. Version 3.0. Fort Myers, Fla: Epi-Systematics Inc, 1984.
  • 7. Q Logic II [computer program]. Version 2.05. Fort Myers, Fla: Epi-Systematics Inc, 1991.
  • 8. EPINET [computer program. Version 1.0.2. Sydney: Becton Dickinson, 1996.
  • 9. SPSS for Windows [computer program]. Version 10.0.7. Chicago, Ill: SPSS Inc, 1999.
  • 10. Epi Info [computer program]. Version 6.0. Atlanta, GA: Centers for Disease Control and Prevention, 1990.
  • 11. Whitby M, Stead P, Najman JM. Needlestick injury: impact of a recapping device and an associated education program. Infect Control Hosp Epidemiol 1991; 12: 220-225.
  • 12. Gershon RRM, Pearse L, Grimes M, et al. The impact of multifocused interventions on sharps injury rates at an acute-care hospital. Infect Control Hosp Epidemiol 1999; 20: 806-811.
  • 13. Mast ST, Woolwine JD, Gerberding JL. Efficacy of gloves in reducing blood volumes transferred during simulated needlestick injury. J Infect Dis 1993; 168: 1589-1592.
  • 14. Goldwater PN, Law R, Nixon AD, et al. Impact of a re-capping device on venepuncture-related needlestick injury. Infect Control Hosp Epidemiol 1989; 10: 21-25.
  • 15. Edmond M, Khakoo R, McTaggart B, et al. Effect of bedside needle disposal units on needle recapping frequency and needlestick injury. Infect Control Hosp Epidemiol 1988; 9: 114-116.
  • 16. Makofksy D, Cone JE. Installing needle disposal boxes closer to the bedside reduces needle-recapping rates in hospital units. Infect Control Hosp Epidemiol 1993; 14: 140-144.
  • 17. Berick DM. A primer on leading the improvement of systems. BMJ 1996; 312: 619-622.
  • 18. Reason J. Human error: models and management. BMJ 2000; 320: 768-770.
  • 19. Jagger J, Hunt EH, Pearson RD. Sharp object injuries in the hospital: causes and strategies for prevention. Am J Infect Control 1990; 18: 227-231.
  • 20. Jagger J, Hunt EH, Pearson RD. Estimated cost of needlestick injuries for six major needled devices. Infect Control Hosp Epidemiol 1990; 11: 584-588.
  • 21. Needlestick Injury Safety and Prevention Act 2000 (US), Section 3, Part 4(A).
  • 22. Younger B, Hunt EH, Robinson C, et al. Impact of a shielded safety syringe on needlestick injuries among healthcare workers. Infect Control Hosp Epidemiol 1992; 13: 349-353.
  • 23. Yassi A, McGill ML, Khokhar JB. Efficacy and cost-effectiveness of a needleless intravenous access system. Am J Infect Control 1995; 23: 57-64.
  • 24. Jagger J, Bentley MB. Injuries from vascular access devices: high risk and preventable. J Intraven Nurs 1997; 20 Suppl: s33-s39.
  • 25. Orenstein R, Reynolds L, Karabaic M, et al. Do protective devices prevent needlestick injuries among healthcare workers? Am J Infect Control 1995; 23: 344-351.
  • 26. L'Ecuyer PB, Schwab EO, Iademarco E, et al. Randomized prospective study of the impact of three needleless intravenous systems on needlestick injury rates. Infect Control Hosp Epidemiol 1996; 17: 803-808.
  • 27. Gerberding JL. Management of occupational exposures to bloodborne viruses. N Engl J Med 1995; 7: 442-451.
  • 28. Patel N, Tignor GH. Device-specific sharps injury and usage rates: an analysis by hospital department. Am J Infect Control 1997; 25: 77-84.
  • 29. Chen LBY, Bailey E, Kogan J, et al. Prevention of needlestick injuries in healthcare workers: 27 month experience with a resheathable safety winged steel needle using CDC NaSH Database [abstract]. Infect Control Hosp Epidemiol 2000; 21: 108.
  • 30. Cockcroft A, Oakley K, Gooch C, et al. Anxiety and perception of risk of HIV and Hepatitis B infection among healthcare workers reporting accidental exposures to blood and other body fluids. AIDS Care 1994; 2: 205-214.
  • 31. Treloar CJ, Malcolm JA, Sutherland DC, et al. Hospital administrators' tolerance of staff needlestick injuries. Infect Control Hosp Epidemiol 1994; 15: 307-310.

Author

remove_circle_outline Delete Author
add_circle_outline Add Author

Comment
Do you have any competing interests to declare? *

I/we agree to assign copyright to the Medical Journal of Australia and agree to the Conditions of publication *
I/we agree to the Terms of use of the Medical Journal of Australia *
Email me when people comment on this article

Online responses are no longer available. Please refer to our instructions for authors page for more information.